Operation instructions and precautions of cm1850 microtome
Author:荣煜化工装备Reader:4679

What are the operating procedures of cm1850 microtome? Generally speaking, it mainly includes the operation instructions and precautions. Therefore, we will focus on the knowledge of these two parts. The specific contents of the operating procedures of cm1850 slicer are as follows.

1. Operation instructions

The operating instructions of cm1850 slicer include two parts: start and slice.

(1) Power on

Turn on the power supply of the host, turn on the switch, and use the box temperature setting button to set the required temperature when slicing. At this time, the temperature display window will synchronously display the set temperature. After setting, after 5 seconds, the actual temperature will be displayed. At the same time, we can also check the set temperature at any time.

When starting the machine for the second time, a reference time should be set through the reference time setting button, and the defrosting time should be set by the defrosting time setting button. If manual defrosting is required, first press the manual defrosting button. After hearing the peak sound, press the quick refrigeration button to defrost the freezing table. Press the box temperature setting button to defrost the box. To close, just press it again in this order.

(2) Slice

The thickness of the slice can be adjusted by using the thickness adjusting knob, and the slice angle should be adjusted when installing the blade.

Before slicing, the sample should be cooled thoroughly, and then put on the slicer, the sample should be close to the knife edge, and the plane should be adjusted. After that, press the button to slice. The cut sample should enter the gap between the anti roll plate and the blade, take it out and observe the staining.

2. Precautions

(1) If you want to prevent the slicer parameters from being changed, you can lock the keyboard through the keyboard lock button on the control panel. Press the button for about 5 seconds. When two points in the time window stop flashing, it means that the keyboard has been locked. The method of unlocking is the same as above.

(2) Before the slicer is turned on again, check whether there is water in the slicer. If so, it should be cleaned before starting up.

(3) Its temperature can be adjusted to below zero for long time if the slicer is used at night.

Unless otherwise specified, the article is original for Rongyu chemical equipment technology, welcome to reprint! Please indicate the address of this article, thank you.